Products

SAM40-16

Sensor Acquisition Module, 16 input channels, 40 kHz bandwidth. Must be used with 12-bit HDO scope.


    SAM40


SAM40-8 - Sensor Acquisition Module, 8 input channels, 40 kHz bandwidth. Must be used with 12-bit HDO.
SAM40-16 - Sensor Acquisition Module, 16 input channels, 40 kHz bandwidth. Must be used with 12-bit HDO.
SAM40-24 - Sensor Acquisition Module, 24 input channels, 40 kHz bandwidth. Must be used with 12-bit HDO.
Key Features
  • 8, 16 or 24 input channels
  • 24-bit resolution
  • 40 kHz bandwidth (100 kS/s)
  • Plug-and-play with supported HDO
  • Complete physical units conversion and rescaling

 

24-bit Precision

All channels have 24-bit resolution (stored in 32-bit floating point format) with ~0.05% total accuracy. Built-in filters with settings as low as 100 Hz further reduce noise.

Plug-and-Play

The SAM40 simply connects to the supported Teledyne LeCroy oscilloscope with a USB2.0 cable for data transfer and control/trigger commands, and BNC cables for clock/timebase synchronization. No programming or complex setups are required.

Unit Conversion and Rescaling

More than 65 different SI and English system physical units are supported for length, mass, temperature, angle, velocity, acceleration, volume, force/weight, pressure, electrical, magnetic, energy and rotating machine quantities.

Deep Toolbox

The oscilloscope math, measure, analysis, pass/fail and option packages utilize the sensor inputs the same as any other channel. Teledyne LeCroy's deep toolbox just got even more powerful.

Key Features
  • 8, 16 or 24 input channels
  • 24-bit resolution
  • 40 kHz bandwidth (100 kS/s)
  • Plug-and-play with supported HDO
  • Complete physical units conversion and rescaling
24-bit Precision

All channels have 24-bit resolution (stored in 32-bit floating point format) with ~0.05% total accuracy. Built-in filters with settings as low as 100 Hz further reduce noise.

Plug-and-Play

The SAM40 simply connects to the supported Teledyne LeCroy oscilloscope with a USB2.0 cable for data transfer and control/trigger commands, and BNC cables for clock/timebase synchronization. No programming or complex setups are required.

Unit Conversion and Rescaling

More than 65 different SI and English system physical units are supported for length, mass, temperature, angle, velocity, acceleration, volume, force/weight, pressure, electrical, magnetic, energy and rotating machine quantities.

Deep Toolbox

The oscilloscope math, measure, analysis, pass/fail and option packages utilize the sensor inputs the same as any other channel. Teledyne LeCroy's deep toolbox just got even more powerful.

Vertical - Analog Channels

Analog Bandwidth (Max) 40 kHz
Bandwidth 40 kHz
Analog Bandwidth @ 1 MΩ (-3 dB) (BNC Input) 40 kHz
Input Channels 16
Input Channels 16
Vertical Resolution 24 bits, stored and displayed as 32-bit floating point waveforms
Vertical Resolution 24 bits
Effective Number of Bits (ENOB) 13 (40 kHz bandwidth), 16 (2 kHz bandwidth)
Vertical Noise Floor (1 mV/div) 31 µVrms
Vertical Noise Floor (2 mV/div) 31 µVrms
Vertical Noise Floor (5 mV/div) 35 µVrms
Vertical Noise Floor (10 mV/div) 35 µVrms
Vertical Noise Floor (20 mV/div) 35 µVrms
Vertical Noise Floor (50 mV/div) 35 µVrms
Vertical Noise Floor (100 mV/div) 35 µVrms
Vertical Noise Floor (200 mV/div) 56 µVrms
Vertical Noise Floor (500 mV/div) 355 µVrms
Vertical Noise Floor (1 V/div) 355 µVrms
Sensitivity 1 mV-10 V/div in 1, 2, 5 steps
DC Vertical Gain Accuracy (Gain Component of DC Accuracy) ±(0.02% of reading + 0.02% F.S. + 100µV) at 23°C ± 5°C. Add ±(0.0015% F.S. + 10µV)/°C outside of stated temperature range.
Channel-Channel Isolation DC-1 kHz: 60 dB (>1000:1),
1-40 kHz: 50 dB (>300:1),
(For any two input channels, same v/div settings, typical)
Maximum Input Voltage ±50 Vpeak for DC, AC, and GND coupling.
±24 Vpeak for IEPE coupling (use IEPE coupling with appropriate sensor).
Input Coupling AC 1MΩ, DC 1MΩ, IEPE, GND.
AC coupling frequency 0.16 Hz ±10%
Input Impedance 1 MΩ ± 0.5% || 55 F
Bandwidth Limiters 100 Hz, 500 Hz, 2 kHz, 10 kHz. Implemented as digital IIR filters in FPGA.
Rescaling Length: meters, inches, feet, yards, miles; Mass: grams, slugs; Temperature: celsius, fahrenheit, kelvin; Angle: radian, arcdegr, arcmin, arcsec, cycles, revolutions, turns; Velocity: m/s, in/s, ft/s, yd/s, miles/s; Acceleration: m/s2, in/s2, ft/s2, g0; Volume: liters, cubic meters, cubic inches, cubic feet, cubic yards; Force (Weight): newton, grain, ounce, pound; Pressure: pascal, bar, atmosphere (technical), atmosphere (standard), torr, psi; Electrical: volts, amps, watts, volt-amperes, volt-amperes reactive, farad, coulomb, ohm, siemen, volt/meter, coulomb/m2, farad/meter, siemen/meter, power factor; Magnetic: weber, tesla, henry, amp/meter, henry/meter; Energy: joule, Btu, calorie; Rotating Machine: radian/second, frequency, revolution/second, revolution/minute, N?m, lb-ft, lb-in, oz-in, watt, horsepower; Other: %.
 

Horizontal - Analog Channels

Acquisition Modes Real-time (coincident with and as controlled by a compatible oscilloscope)
Clock Accuracy +/-25ppm + 1.0ppm/year from calibration
Jitter Between Channels SAM40-SAM40 Channel: 1µs (maximum)
Oscilloscope Analog Channel-SAM40 Channel: 10µs (maximum)
 

Acquisition - Analog Channels

Sample Rate 100 kS/s on all 8 channels
Maximum Acquisition Memory 2.5 Mpts on all 8 channels
Sample Rate (Single-shot) 1 S/s to 100 kS/s on all channels
Memory Length Options (4 Ch / 2 Ch / 1Ch)
(Number of Segments)
2.5 Mpts/ch on all channels
Averaging Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps
Envelope (Extrema) Envelope, floor, or roof for up to 1 million sweeps
 

Triggering System

Modes Normal, Auto, Single, and Stop when cross-triggered by a supported oscilloscope.
 

Connectivity

Ethernet Port Qty. 1 10/100/1000BaseT Ethernet interface (RJ45 port).
USB Device Port 1 USBTMC port (USB 2.0)
 

Probes

Probe System BNC Input, IEPE/ICP Compatible. Excitation Output Current = 4mA ±15% (23°C ± 5°C, +0.5%/°C outside of range). Compliance Voltage = 24V.
Scale Factors Manually selected in supported oscilloscope.
 

Power Requirements

Voltage 100-240 VAC ±10% at 45-66 Hz; 100-120 VAC ±10% at 380-420 Hz; Automatic AC Voltage Selection.
Nominal Power Consumption 35 W / 35 VA
Max. Power Consumption 50 W / 50 VA
 

Environmental

Temperature (Operating) 0 °C to +50 °C
Temperature (Non-Operating) -40 °C to +70 °C
Humidity (Operating) 5% to 90% relative humidity (non-condensing) up to +30 °C.
Upper limit derates to 50% relative humidity (non-condensing) at +50 °C.
Humidity (Non-Operating) 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F
Altitude (Operating) Up to 10,000 ft. (3048 m) at or below +25 °C
Altitude (Non-Operating) Up to 40,000 ft. (12,192 m)
Random Vibration (Operating) 0.31 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Random Vibration (Non-Operating) 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Functional Shock 30 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total
 

Size and Weight

Dimensions (HWD) 3.97"H x 16.79"W x 13.38"D (101 mm x 426 mm x 340 mm)
Weight 9.56 lbs (4.35 kg)
 

Certifications

CE Certification
UL and cUL Listing
CE Compliant
 

Warranty and Service

Warranty and Service 3-year warranty; calibration recommended annually.
Optional service programs include extended warranty, upgrades, and calibration services.
 

SAM40-16