LabMaster 10 Zi-A 50 GHz Systems

50 GHz, 160 GS/s, 64 Mpts/Ch Digital Bandwidth Interleaved Acquisition Module, 50 Ohm Input Also operates in 36 GHz, 80 GS/s, 4ch, 32 Mpts/Ch mode.


  • product line tab
  • specs tab

    LabMaster Acquisition Modules


LabMaster 10 Zi-A Oscilloscopes
LabMaster 10-20Zi-A - 20 GHz, 80 GS/s, 4ch, 32 Mpts/Ch LabMaster Acquisition Module, 50 Ohm Input
LabMaster 10-25Zi-A - 25 GHz, 80 GS/s, 4ch, 32 Mpts/Ch LabMaster Acquisition Module, 50 Ohm Input
LabMaster 10-30Zi-A - 30 GHz, 80 GS/s, 4ch, 32 Mpts/Ch LabMaster Acquisition Module, 50 Ohm Input
LabMaster 10-36Zi-A - 36 GHz, 80 GS/s, 4ch, 32 Mpts/Ch LabMaster Slave Acquisition Module, 50 Ohm Input
LabMaster 10-50Zi-A - 50 GHz, 160 GS/s, 64 Mpts/Ch Digital Bandwidth Interleaved Acquisition Module, 50 Ohm Input Also operates in 36 GHz, 80 GS/s, 4ch, 32 Mpts/Ch mode. Limited Availability
LabMaster 10-59Zi-A - 59 GHz, 160 GS/s, 64 Mpts/Ch Digital Bandwidth Interleaved Acquisition Module, 50 Ohm Input Also operates in 36 GHz, 80 GS/s, 4ch, 32 Mpts/Ch mode. Limited Availability
LabMaster 10-65Zi-A - 65 GHz, 160 GS/s, 64 Mpts/Ch Digital Bandwidth Interleaved Acquisition Module, 50 Ohm Input. Also operates in 36 GHz, 80 GS/s, 4ch, 32 Mpts/Ch mode. Limited Availability

Vertical - Analog Channels

Analog Bandwidth (Max) 50 GHz
Bandwidth 50 GHz
Analog Bandwidth @ 50 Ω (-3 dB) (1.85mm Inputs) 50 GHz
(≥10 mV/div)
Analog Bandwidth @ 50 Ω (-3 dB) (2.4/2.92mm Inputs) 36 GHz
(≥5 mV/div)
Rise Time (10-90%, 50 Ω) 8 ps
(test limit, flatness mode)
Rise Time (20-80%, 50 Ω) 6 ps
(flatness mode)
Input Channels Up to 40, depending on configuration selected.
Up to 80 @ 36 GHz.
Input Channels Up to 40 @ 50 GHz,
Up to 80 @ 36 GHz
Vertical Resolution 8-bits
Vertical Resolution 8 bits; up to 11 bits with enhanced resolution (ERES)
Sensitivity 50 Ω (2.92mm): 5 mV-500mV/div, fully variable (5-9.9 mV/div via zoom)
50 Ω (1.85mm): 10 mV-80mV/div, fully variable. Higher gain settings possible through use of external attenuators.
DC Vertical Gain Accuracy (Gain Component of DC Accuracy) ±1% F.S. (typical), offset at 0V; ±1.5% F.S. (test limit), offset at 0V
Channel-Channel Isolation DC to 36 GHz: 60 dB (>1000:1)
(For any two 2.92mm input channels, same or different v/div settings, typical)
36 to 65 GHz: 40 dB (>100:1)
(For any two 1.85mm input channels, same or different v/div settings, typical)
Offset Range 50 Ω (1.85 mm):
±500 mV @ 10-80 mV/div
50 Ω (2.92mm):
±500 mV @ 5-75 mV/div
±4 V @ 76 mV/div -500mV/div
DC Vertical Offset Accuracy ±(1.5% of offset setting + 1.5% F.S. + 1 mV) (test limit)
Maximum Input Voltage 2.92 mm Inputs: ±2 Vmax@<76mV/div, 5.5Vrms@≥76mV/div
1.85 mm Inputs: ±2 Vmax@≤80mV/div
Input Coupling 2.92 mm Inputs: 50 Ω: DC, GND
1.85 mm Inputs: 50 Ω: DC
Input Impedance 2.92mm Inputs: 50 Ω+/-2%
1.85mm Inputs: 50 Ω+/-2%
Bandwidth Limiters For ≤ 36 GHz Mode: 1, 3, 4, 6, 8, 13, 16, 20, 25, 30, 33 GHz.
For > 36 GHz Mode: None.
Rescaling Length: meters, inches, feet, yards, miles; Mass: grams, slugs; Temperature: celsius, fahrenheit, kelvin; Angle: radian, arcdegr, arcmin, arcsec, cycles, revolutions, turns; Velocity: m/s, in/s, ft/s, yd/s, miles/s; Acceleration: m/s2, in/s2, ft/s2, g0; Volume: liters, cubic meters, cubic inches, cubic feet, cubic yards; Force (Weight): newton, grain, ounce, pound; Pressure: pascal, bar, atmosphere (technical), atmosphere (standard), torr, psi; Electrical: volts, amps, watts, volt-amperes, volt-amperes reactive, farad, coulomb, ohm, siemen, volt/meter, coulomb/m2, farad/meter, siemen/meter, power factor; Magnetic: weber, tesla, henry, amp/meter, henry/meter; Energy: joule, Btu, calorie; Rotating Machine: radian/second, frequency, revolution/second, revolution/minute, N?m, lb-ft, lb-in, oz-in, watt, horsepower; Other: %.
 

Horizontal - Analog Channels

Timebases Internal timebase with 10 GHz clock frequency common to all input channels. Single, distributed 10 GHz clock for all channels ensures precise synchronization with timing accuracy between all channels identical to that provided within a single, conventional oscilloscope package.
Time/Division Range For >36 GHz Mode: 10 ps/div - 320 µs/div (maximum capture time is based on 160 GS/s and installed memory).
For ≤36 GHz Mode: 10 ps/div-256 s/div (maximum capture time is based on minimum sample rate of 200kS/s and installed memory).
Clock Accuracy <0.1 ppm + (aging of 0.05 ppm/yr from last calibration)
Sample Clock Jitter up to 3.2ms Acquired Time Range:
50fsrms (Internal Timebase Reference)
50fsrms (External Timebase Reference)
up to 6.4ms Acquired Time Range:
130fsrms (Internal Timebase Reference)
130fsrms (External Timebase Reference)
Delta Time Measurement Accuracy \sqrt{2}*\sqrt{(\frac{Noise}{SlewRate})^2+(Sample Clock Jitter)^2) (RMS)+(clock accuracy*reading)(seconds)
Jitter Measurement Floor \sqrt{(\frac{Noise}{SlewRate})^2+(Sample Clock Jitter)^2\; seconds\; rms\; (TIE)}
Jitter Between Channels <190fsrms
(TIE, typical, measured at maximum bandwidth)
Channel-Channel Deskew Range ±9 x time/div. setting or 25 ns max. (whichever is larger), each channel
External Timebase Reference (Input) 10 MHz or 100MHz; 50 Ω impedance, applied at the rear input of MCM-Zi-A Master Control Module
External Timebase Reference (Output) 10 MHz or 100 MHz ; 50 Ω impedance, output at the rear of MCM-Zi-A Master Control Module
 

Acquisition - Analog Channels

Sample Rate 80 GS/s on 4 Ch
160 GS/s on 2 Ch
Maximum Acquisition Memory 1024 Mpts/Ch (2 Ch operation)
Sample Rate (Single-shot) 160 GS/s on each channel in >36 GHz Mode.
80 GS/s on each channel in ≤36 GHz Mode.
Memory Length (4 Ch / 2 Ch / 1Ch)
(Number of Segments)
64 M / 64 M / 64 M
(In ≤36 GHz Modes, reference memory specification for 36 GHz LabMasters)
(1000)
Memory Length Options (4 Ch / 2 Ch / 1Ch)
(Number of Segments)
S-32 Option:
64M / 64M / 64M
(3,500)
M-64 Option:
128M / 128M / 128M
(7,500)
L-128 Option:
256M / 256M / 256M
(15,000)
VL-256 Option:
512M / 512M / 512M
(15,000)
XL-512 Option:
1024M / 1024M / 1024M
(15,000)
(In ≤36 GHz Modes, reference memory specification for LabMaster 10 Zi-A 36 GHz Systems)
Intersegment Time 1 µs
Averaging Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps
Enhanced Resolution (ERES) From 8.5 to 11 bits vertical resolution
Envelope (Extrema) Envelope, floor, or roof for up to 1 million sweeps
Interpolation Linear or Sin x/x
 

Triggering System

Modes Normal, Auto, Single, and Stop
Sources Any Ch 1-4 (Edge, Window, SMART, Cascade triggers), AUX, internal Fast Edge; or any input channel (Edge trigger only) on additional 10-xxZi-A Acquisition Modules (Channels 5 and higher).
Slope and level unique to each source except line trigger.
Coupling DC, AC, HFRej, LFRej
Pre-trigger Delay 0-100% of memory size (adjustable in 1% increments of 100 ns)
Post-trigger Delay 0-10,000 divisions in real time mode, limited at slower time/div settings
Hold-off From 2 ns up to 20 s or from 1 to 99,999,999 events
Trigger and Interpolator Jitter <0.1 ps rms (typical, software assisted), 2 ps rms (typical, hardware),
Internal Trigger Level Range ±4.1 div from center
External Trigger Level Range For any LabMaster 10xx-Zi-A Acquisition Module: Aux (±0.4 V)
(Only Ch1-4 Acquisition Module has "active" AUX Input)
Maximum Trigger Rate 1,000,000 waveforms/second (in Sequence Mode, up to 4 channels)
Trigger Sensitivity with Edge Trigger (1.85/2.4/2.92mm Inputs) For Ch 1-80 of a LabMaster 10 Zi-A system:
3 div @ < 12 GHz
1.5 div @ < 3 GHz
1.0 div @ < 200 MHz
(for DC coupling, ≥ 10 mV/div, 50 Ω )
Trigger Sensitivity with Edge Trigger (Aux Input) For Ch 1-4 only of any LabMaster 10xx-Zi-A Acquisition Module:
2 div @ < 1 GHz,
1.5 div @ < 500 MHz,
1.0 div @ < 200 MHz,
(for DC coupling)
Max. Trigger Frequency, SMART Trigger For Ch 1-4 only of any LabMaster 10xx-Zi-A Acquisition Module:
2.0 GHz @ ≥ 10 mV/div (minimum triggerable width 200 ps)
 

Trigger Types

Edge Triggers when signal meets slope (positive, negative, or either) and level condition.
Width Triggers on positive, negative, or both widths with widths selectable as low as 200ps to 20 s, or on intermittent faults.
Glitch Triggers on positive or negative glitches with widths selectable as low as 200ps to 20 s, or on intermittent faults.
Window Triggers when signal exits a window defined by adjustable thresholds
Pattern Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input). Each source can be high, low, or don’t care. The High and Low level can be selected independently. Triggers at start or end of the pattern.
Runt Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1 ns and 20 ns.
Slew Rate Trigger on edge rates. Select limits for dV, dt, and slope. Select edge limits between 1 ns and 20 ns.
Interval Triggers on intervals selectable between 1 ns and 20 s.
Dropout Triggers if signal drops out for longer than selected time between 1 ns and 20 s.
Measurement Trigger Select from a large number of measurement parameters trigger on a measurement value with qualified limits. Can be used as only trigger or last event in a Cascade Trigger.
Multi-Stage: Qualified (Timeout or State/Edge Qualified) Triggers on any input source only if a defined state or edge occurred on another input source. Holdoff between sources is selectable by time or events.
Triggers with Exclusion Technology Trigger on intermittent faults by specifying the expected behavior and triggering when that condition is not met
Multi-Stage: Qualified First In Sequence acquisition mode, triggers repeatably on event B only if a defined pattern, state, or edge (event A) is satisfied in the first segment of the acquisition. Holdoff between sources is selectable by time or events.
Multi-Stage: Cascade (Sequence) Trigger, Capability Arm on "A" event, then Trigger on "B" event. Or Arm on "A" event, then Qualify on "B" event, and Trigger on "C" event.
Multi-Stage: Cascade (Sequence) Trigger, Types Cascade A then B: Edge, Window, Pattern (Logic) Width, Glitch, Interval, Dropout, or Measurement. Measurement can be on Stage B only.
Cascade A then B then C (Measurement): Edge, Window, Pattern (Logic), Width, Glitch, Interval, Dropout, or Measurement. Measurement can be on Stage C only.
Cascade A then B then C: Edge, Window, Pattern (Logic)
Multi-Stage: Cascade (Sequence) Trigger, Holdoff Holdoff between A and B or B and C is selectable by time or number of events.
Measurement trigger selection as the last stage in a Cascade precludes a holdoff setting between the prior stage and the last stage.
 

High Speed Serial Protocol Trigger

Data Rates Option LM10Zi-6GBIT-80B-SYMBOL-TD: 600 Mb/s to 6.5 Gb/s, Channel 4 input only
Option LM10Zi-14GBIT-80B-SYMBOL-TD: 600 Mb/s to 14.1 Gb/s, Channel 4 input only
(Note: Channel 3 input will capture signal for triggering when oscilloscope is in ≥25 GHz mode)
Pattern Length 80 bits NRZ, eight 8b/10b symbols, 64b/66b symbol
 

Display System

Size On LabMaster MCM-Zi Master Control Module: Color 15.3" flat panel TFT-Active Matrix LCD with high resolution touch screen
Resolution WXGA; 1280 x 768 pixels.
Number of traces Display a maximum of 40 traces. Simultaneously display channel, zoom, memory and math traces.
Grid Styles Auto, Single, Dual, Triple, Quad, Octal, X-Y, Single+X-Y, Dual+X-Y, Twelve, Sixteen. Up to twenty grids available with some software options.
Waveform Representation Sample dots joined, or sample dots only
 

Integrated Second Display

Size Supports touch screen integration of user-supplied second display with split-grid capability.
(Note: touch screen driver for second display may not be a Fujitsu driver)
Resolution Determined by display chosen by user
 

LeCroy WaveStream™ Fast Viewing Mode

Operation Front panel toggle between WaveStream ON (Analog), ON (Color) and OFF
Intensity 256 Intensity Levels, 1-100% adjustable via front panel control
Types Select analog or color-graded
Number of Channels up to 80 simultaneously
Max Sampling Rate 160GS/s
Persistence Aging Select from 500 ms to Infinity
Waveforms/second (continuous) up to 2500 Waveforms/second
 

Persistence Display

Analog and Color-Graded Persistence Variable saturation levels; stores each trace’s persistence data in memory
Persistence Types Select analog, color, or three-dimensional
Trace Selection Activate persistence on all or any combination of traces
Persistence Aging Select from 500 ms to infinity
Sweep Display Modes All accumulated, or all accumulated with last trace highlighted
 

High Speed Digitizer Output (Option)

Type Option LSIB-2. Installs in LabMaster MCM-Zi Master Control Module and uses one available PCIe slot normally used by a LabMaster 10-xxZi Acquisition Module.
Transfer Rate up to 325 MB/s (typical) - Maximum of 4 channels (consult Teledyne LeCroy for >4 channels)
Output Protocol PCI Express, Gen 1 (4 lanes utilized for data transfer)
Control Protocol TCP/IP
Command Set Via Windows Automation, or via LeCroy Remote Command Set
 

Zoom Expansion Traces

Zoom Expansion Traces Display up to 12 Zoom and 12 Math/Zoom traces
 

Internal Waveform Memory

Internal Waveform Memory 12 active waveform memory traces (M1-M12) store 16 bit/point full length waveforms. Waveforms can be stored to any number of files limited only by the data storage media capacity.
 

Setup Storage

Front Panel and Instrument Status Store to the internal hard drive, over the network, or to a USB-connected peripheral device.
 

Automatic Setup

Auto Setup Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals
Find Vertical Scale Automatically sets the vertical sensitivity and offset for the selected channel to display a waveform with the maximum dynamic range
 

Calibration

Auto Calibration Ensures specified DC and timing accuracy is maintained for 1 year minimum.
 

Processor/CPU

Processor Memory 32 GB standard. Up to 192 GB optionally available.
Operating System Microsoft Windows® 10
Oscilloscope Operating Software Teledyne LeCroy MAUI™ with OneTouch
Real Time Clock Date and time displayed with waveform in hardcopy files. SNTP support to synchronize to precision internal clocks.
 

Connectivity

Ethernet Port Supports 10/100/1000BaseT Ethernet interface (RJ45 port)
USB Host Ports LabMaster MCM-Zi -A Master Control Module: minimum 2 total USB 2.0 ports on rear of unit to support Windows compatible devices
LabMaster MCM-Zi -A Master Control Module: minimum 3 total USB 2.0 ports on front of unit to support Windows compatible devices
GPIB Port (optional) Supports IEEE - 488.2. Installs in LabMaster MCM-Zi Master Control Module and uses one available PCIe slot normally used by a LabMaster 10-xxZi Acquisition Module.
External Monitor Port Dual Link DVI compatible to support internal display on MCM-Zi Master Control Module (1280 x 768 pixel resolution) and customer-supplied monitor with up to WQXGA (2560 x 1600 pixel) resolution using extended desktop mode.
Remote Control Via Windows Automation, or via LeCroy Remote Command Set
Network Communication Standard VXI-11 or VICP, LXI Class C (v1.2) Compliant
LSIB Port (optional) Supports PCIe Gen1 x4 protocol with LeCroy supplied API. Installs in LabMaster LabMaster MCM-Zi-A Master Control Module and uses one available PCIe slot normally used by a LabMaster 10-xxZi-A Acquisition Module.
 

Auxiliary Input

Signal Types For External Trigger Input. Located on every LabMaster 10-xxZi-A Acquisition Module (front-mounted BNC, 50 Ω). 10-xxZi-A Acquisition Module that is used for Channels 1-4 contains "active" AUX Input.
Coupling 50 Ω: DC
Max. Input Voltage 5 Vrms
 

Auxiliary Output

Signal Types Select from calibrator, control signals or Off
Output Signal 500 Hz-5 MHz square wave or DC level; 0.0 to 500 mV into 50 Ω (0-1 V into 1 MΩ)
Control Signals Trigger enabled, trigger out, pass/fail status, off
Connector Type BNC, located on front of every LabMaster 10-xxZi Acquisition Module. 10-xxZi Acquisition Module that is used for Channels 1-4 contains "active" AUX Output.
 

Probes

Probes No passive probes supplied
Probe System Up to 36 GHz Inputs: 2.92mm interface.
Up to 65 GHz Inputs: 1.85mm interface
2.92mm interface automatically detects and supports a variety of compatible probes.
Use L2.92A-PLINK or L2.92A-PBUS 2.92mm Probe Adapters to connect 50Ω ProLink or ProBus compatible probes to 2.92mm interface.
Scale Factors Automatically or manually selected depending on probe used
Calibration Output MCM-Zi-A Master Control Module: Fast Edge output, 5MHz square wave, 350mVp-p (typical), 50ps rise time, output to SMA connector, AC Coupled
 

Power Requirements

Voltage LabMaster 10-xxZi Acquisition Module: 100-240 VAC ±10% at 45-66 Hz; 100-120 VAC ±10% at 380-420 Hz; Automatic AC Voltage Selection, Installation Category II
LabMaster MCM-Zi Master Control Module: 100-240 VAC ±10% at 45-66 Hz; Automatic AC Voltage Selection, Installation Category II
Max. Power Consumption LabMaster 10-xxZi-A Acquisition Module - 1275 W / 1275 VA.
LabMaster MCM-Zi-A Master Control Module - 450 W / 450 VA.
Each Module and the CPU has a separate power cord.
 

Certifications

Certifications CE Compliant, UL and cUL listed; conforms to EN 61326, EN 61010-1, EN61010-2-030, UL 61010-1 3rd edition, and CSA C22.2 No. 61010-1-12
 

Environmental

Temperature (Operating) +5 °C to +40 °C
Temperature (Non-Operating) -20 °C to +60 °C
Humidity (Operating) 5% to 80% relative humidity (non-condensing) up to +31 °C.
Upper limit derates to 50% relative humidity (non-condensing) at +40 °C.
Humidity (Non-Operating) 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F
Altitude (Operating) Up to 10,000 ft. (3048 m) at or below +25 °C
Altitude (Non-Operating) Up to 40,000 ft. (12,192 m)
Random Vibration (Operating) 0.5 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Random Vibration (Non-Operating) 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Functional Shock 20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total
 

Size and Weight

Dimensions (HWD) LabMaster MCM-Zi-A Master Control Module - 10.9"H x 18.2"W x 15.6"D (277 x 462 x 396 mm),
LabMaster 10-xxZi-A Acquisition Module - 8.0"H x 18.2"W x 26"D (202 x 462 x 660 mm)
Weight LabMaster 10-xxZi-A Acquisition Module -58 lbs. (26.4 kg)
LabMaster MCM-Zi-A Master Control Module - 47 lbs. (21.4 kg),
Shipping Weight LabMaster 10-xxZi-A Acquisition Module -76 lbs. (34.5 kg)
LabMaster MCM-Zi-A Master Control Module - 56 lbs. (25.5 kg)
 

Warranty and Service

Warranty and Service 3-year warranty; calibration recommended annually.
Optional service programs include extended warranty, upgrades, and calibration services.
 