VOLTAGE CONTROLą3.5 kV Configurationą7 kV Configuration
Range100-3500 V200-7000 V
CAMAC Programming14 bits14 bits
CAMAC Step0.25 V0.5 V
CAMAC Accuracyē ą(1.5 V + 0.1%)ē ą(3 V + 0.1%)
Front-Panel AdjustmentTen-turn vernier potentiometer
500 V/turn1000 V/turn
CAMAC Monitor12-bit ADC12-bit ADC
CAMAC Resolution1 V2 V
CAMAC Offset (response to 0 V output)0-2 counts0-2 counts
CAMAC Accuracyą0.2%ą0.2%
Front-Panel MonitorLow Impedance (< 10 ohm) voltage output, suitable for driving a meter; BNC connector.
Front-Panel Scale2 V/kV ą0.5%1 V/kV ą0.5%
Front-Panel Offset (value for 0 V output)ē ą4 V referred to HV outputē ą8 V referred to HV output
CURRENT CONTROL
Range0-2.5 mA0-1 mA
CAMAC Programming8 bits8 bits
CAMAC Step10 ĩA4 ĩA
CAMAC Accuracyą(1% + 12 ĩA)ą(1% + 5 ĩA)
Front-Panel AdjustmentTen-turn vernier potentiometer
250 ĩA/turn100 ĩA/turn
CAMAC Monitor12-bit ADC12-bit ADC
CAMAC Resolution625 nA250 nA
CAMAC Offset (Response to 0 ĩA load)1 or 2 counts1 or 2 counts
CAMAC Accuracyą0.5%ą0.5%
Front-Panel MonitorLow Impedance (< 10 ohm) voltage output, suitable for driving a meter; BNC connector.
Front-Panel Scale4 V/mA ą0.5%10 V/mA ą0.5%
Front-Panel Offset (value for 0 V output)ē ą4 ĩAē ą2 ĩA
GENERAL
Output Power Rating7 W7 W
Output Ripple (at maximum current)< 50 mV rms< 50 mV rms
Output Temperature CoefficientTypically ą50 ppm/ĄC, ą100 ppm/ĄC maximum
Voltage Regulation< 0.5 V (0 to 2.5 mA)< 1 V (0 to 1 mA)
Short Circuit ProtectionYesYes
Output ConnectorSHVReynolds 1064-1
CONTROLS
HV On/OffToggle switch; turns On/Off HV output.
Manual/RemoteSelects front panel or CAMAC operation of voltage and current set.
ADC12 bits; conversion time < 30 ĩsec.
INDICATOR LAMPS
HV OnVoltage present at output.
OverloadSupply in current-limiting mode.
PositivePositive output configuration.
NegativeNegative output configuration.
3.5 kVONOFF
7 kVOFFON
POWER350 mA at +24 V; 350 mA at -24 V; 350 mA at +6 V.