The bandwidth required for capture and measurement of signals
                                                            depends greatly on the signals to be measured, the types of
                                                            measurements to be made, and the accuracy desired of the
                                                            measurements. A rough rule of thumb most engineers use is to
                                                            have an oscilloscope with three times the bandwidth of the
                                                            highest frequency signal they wish to measure, though this
                                                            becomes impractical for very high frequency signals.
                                                        Reference the definition for oscilloscope bandwidth in the
                                                            FAQ (above). Most oscilloscopes approach the -3 dB
                                                            bandwidth-rated frequency slowly, beginning with a gentle
                                                            amplitude rolloff at 50% (or so) of the bandwidth frequency
                                                            rating. This means that if the oscilloscope amplitude
                                                            response is -1 dB at 70% of rated bandwidth and -2 dB at 85%
                                                            of rated bandwidth, then the amplitude of the captured pure
                                                            sinusoid will be approximately 90% (-1 dB) or 80% (-2 dB)
                                                            and 70% (-3 dB) compared to when the input sinusoid
                                                            frequency is approaching the bandwidth rating of the
                                                            oscilloscope. However, most engineers are not measuring pure
                                                            sinusoids with their oscilloscope. Note that the highest
                                                            bandwidth oscilloscopes may have a flatter (less amplitude
                                                            rolloff) or adjustable amplitude response, for a variety of
                                                            reasons.
                                                        More likely, an engineer is measuring a signal that resembles
                                                            a square wave. In this case, it is known that a square wave
                                                            can be represented as a Fourier series expansion comprised
                                                            of the sum of the fundamental frequency and odd harmonics,
                                                            with the Nth harmonic contributing a 1/N amplitude at that
                                                            frequency. What this means is that to accurately represent a
                                                            square wave, you need enough bandwidth to capture the
                                                            fundamental frequency and enough of the odd harmonics. How
                                                            many odd harmonics is “enough” (and how much
                                                            bandwidth is needed) is determined by the engineer’s
                                                            tolerance for a rise time measurement on the oscilloscope
                                                            that is slower than the real signal, and the amount of
                                                            additive overshoot and ringing present on the measured
                                                            signal. If only the 3rd harmonic is captured, the rise time
                                                            will be appreciably slower, and the overshoot and ringing
                                                            will be noticeable compared to if the 99th harmonic is
                                                            captured (in which case the captured signal will be
                                                            indistinguishable from the original input signal).
                                                        This gets us back to the original answer that is given most
                                                            often in response to the question of “how much
                                                            bandwidth is needed?” – about 3x the bandwidth
                                                            of the highest frequency signal. But what does
                                                            “highest frequency” mean? In this context, most
                                                            engineers are thinking of the rise time measurement
                                                            capability of the oscilloscope (which is related to
                                                            bandwidth). If an engineer wants to measure a signal with a
                                                            rise time of 1 ns, they would not choose an oscilloscope
                                                            with a 1 ns rise time (such an oscilloscope would typically
                                                            have a bandwidth of 350 MHz) – they would choose an
                                                            oscilloscope with bandwidth 3x that (or 1 GHz).
                                                        Reference webinarPart
                                                                2: How Much Bandwidth Do I Need in My
                                                                Oscilloscope?in the 2023 Oscilloscope Coffee Break
                                                            Webinar Series for other details.