There are often instances in which an oscilloscope user may need to capture a large number of fast pulses in quick succession, or, conversely, a small number of events separated by long periods of time. Either scenario can be challenging with typical oscilloscope acquisition modes. Sequence Mode, a standard acquisition mode found in the HDO4000 and HDO6000 High Definition Oscilloscopes, enables capture of fine details of complex event sequences occurring over long time intervals, while ignoring the intervals between events.
For this demonstration, the input signal on channel 1 is from an echo ranging system. A transmitted pulse is followed by a reflected pulse whose position and amplitude vary with distance. The measurement repeats at a 10 kHz (100 µs) rate. The initial oscilloscope setup is shown in Figure 1.
Sequence Mode can be accessed by tapping the Timebase descriptor box or by opening the Timebase pulldown menu and clicking Horizontal Setup. On the Timebase tab, select Sequence under Sampling Mode. This will create a new tab in the Timebase dialog box labeled Sequence.
The first section on the left side of the Sequence tab is Acquisition Settings. Touching the Number of Segments button twice brings up a dialog box that allows setting of how many acquisitions (or segments) the oscilloscope will make in Sequence mode. This can range from a minimum of two to a maximum of 10,000 with the HDO4000 and 65,000 with the HDO6000; the default number is 10. Regardless of what number is entered, the oscilloscope will acquire data until either that number is reached, the timeout (described next) is reached, or the user presses the front-panel Stop button. Next to the Acquisition Settings section is the Sequence Timeout section. Here, the user may instruct the oscilloscope to stop acquiring data if a preset time period passes without a trigger event.
When using the HDO4000, segments are shown using an adjacent display, with each segment shown next to the previous segment. For long captures of many segments, the Zoom function can be used to show individual segments. The HDO6000 offers a variety of segment-viewing techniques. The Display Settings section has all the display modes, which include adjacent, overlay, waterfall, perspective, and mosaic. The adjacent mode is the default as seen in Figure 2. Another option, Perspective mode, is shown in Figure 3.
Also within the Display Settings section of the Sequence tab is the option to Show Sequence Trigger Times. Tapping this button brings up an entirely new Trigger Times dialog box shown in Figure 4.
The first column under the Time section is the segment number, followed by an absolute time stamp (1s resolution). The third column shows the time since Segment #1 and the fourth shows the time between segments (both in nanosecond resolution).
Each segment can be viewed individually or as groups of segments by pushing the front-panel Zoom button or by using the touch screen, as shown in Figure 5. The horizontal delay knob then permits scrolling through the segments. In the Zoom dialog box, users can set up zoomed views of multiple segments for comparison purposes.
The Sequence Mode acquisition mode found in the HDO4000 and HDO6000 is a fast trigger mode that makes the most of the instruments’ acquisition memory. With each memory segment storing a single acquisition, Sequence Mode minimizes dead time between trigger events and enables capture of many fast pulses in rapid succession.