Vertical - Analog Channels |
Analog Bandwidth (Max) | 2 GHz |
Bandwidth | 2 GHz |
Analog Bandwidth @ 50 Ω (-3 dB) (ProBus Input) | 2 GHz (≥1 mV/div) |
Analog Bandwidth @ 1 MΩ (-3 dB) (ProBus Input) | 500 MHz |
Rise Time (10-90%, 50 Ω) | 235 ps |
Rise Time (20-80%, 50 Ω) | 176 ps |
Input Channels | 8 |
Vertical Resolution | 12 bits; up to 15 bits with enhanced resolution (ERES) |
Effective Number of Bits (ENOB) | 8.4 bits |
Vertical Noise Floor (1 mV/div) | 170 µVrms |
Vertical Noise Floor (2 mV/div) | 170 µVrms |
Vertical Noise Floor (5 mV/div) | 175 µVrms |
Vertical Noise Floor (10 mV/div) | 200 µVrms |
Vertical Noise Floor (20 mV/div) | 235 µVrms |
Vertical Noise Floor (50 mV/div) | 330 µVrms |
Vertical Noise Floor (100 mV/div) | 510 µVrms |
Vertical Noise Floor (200 mV/div) | 2.35 mVrms |
Vertical Noise Floor (500 mV/div) | 3.25 mVrms |
Vertical Noise Floor (1 V/div) | 5.20 mVrms |
Sensitivity | 50 Ω: 1 mV-1 V/div, fully variable; 1 MΩ: 1 mV-10 V/div, fully variable |
DC Vertical Gain Accuracy (Gain Component of DC Accuracy) | ±(0.5%) FS, offset at 0 V |
Channel-Channel Isolation | DC-200 MHz: 70 dB (>3000:1), 200-500 MHz: 60 dB (>1000:1), 500 MHz - 1 GHz: 50 dB (>300:1), 1 GHz - 2 GHz: 40 dB (>100:1) (For any two input channels, same V/div settings, typical) |
Offset Range | 50 Ω ±1.6 V @ 1 mV - 4.95 mV/div ±4 V @ 5 mV - 9.9 mV/div ±8 V @ 10 mV - 19.8 mV/div ±10 V @ 20 mV - 1 V/div
1 MΩ ±1.6 V @ 1 mV - 4.95 mV/div ±4 V @ 5 mV - 9.9 mV/div ±8 V @ 10 mV - 19.8 mV/div ±16 V @ 20 mV - 100 mV/div ±80 V @ 102 mV - 198 mV/div ±160 V @ 200 mV - 1 V/div ±400 V @ 1.02 V - 10 V/div |
DC Vertical Offset Accuracy | ±(0.5% of offset setting + 0.5% FS + 1mV) |
Maximum Input Voltage | 50 Ω: 5 Vrms, 1 MΩ: 400 V max (DC + Peak AC ≤ 10 kHz) |
Input Coupling | 1 MΩ: AC, DC, GND; 50 Ω: DC, GND |
Input Impedance | 50 Ω+/-2% or 1 MΩ||19pF, 10 MΩ || 10 pF |
Bandwidth Limiters | 20MHz, 200MHz, 350 MHz, 500 MHz, 1 GHz |
Rescaling | Length: meters, inches, feet, yards, miles; Mass: grams, slugs; Temperature: celsius, fahrenheit, kelvin; Angle: radian, arcdegr, arcmin, arcsec, cycles, revolutions, turns; Velocity: m/s, in/s, ft/s, yd/s, miles/s; Acceleration: m/s2, in/s2, ft/s2, g0; Volume: liters, cubic meters, cubic inches, cubic feet, cubic yards; Force (Weight): newton, grain, ounce, pound; Pressure: pascal, bar, atmosphere (technical), atmosphere (standard), torr, psi; Electrical: volts, amps, watts, volt-amperes, volt-amperes reactive, farad, coulomb, ohm, siemen, volt/meter, coulomb/m2, farad/meter, siemen/meter, power factor; Magnetic: weber, tesla, henry, amp/meter, henry/meter; Energy: joule, Btu, calorie; Rotating Machine: radian/second, frequency, revolution/second, revolution/minute, N m, lb-ft, lb-in, oz-in, watt, horsepower; Other: %. |
|
Horizontal - Analog Channels |
Timebases | Internal timebase common to 8 input channels |
Acquisition Modes | Real-time, Roll, Sequence |
Time/Division Range | 100 ps/div - 5 ks/div with standard memory (up to 10 ks/div with 500MPT memory, 25 ks/div with 1000MPT memory, 50 ks/div with 2000MPT memory, 100 ks/div with 5000MPT memory); Roll Mode available at ≥ 100 ms/div and ≤ 5 MS/s |
Clock Accuracy | +/-1 ppm + 1 ppm/year from calibration |
Sample Clock Jitter | up to 10µs Acquired Time Range: 80 fsrms (Internal Timebase Reference)
up to 10ms Acquired Time Range: 150 fsrms (Internal Timebase Reference) |
Delta Time Measurement Accuracy | \sqrt{2}*\sqrt{(\frac{Noise}{SlewRate})^2+(Sample Clock Jitter)^2) (RMS)+(clock accuracy*reading)(seconds) |
Jitter Measurement Floor | \sqrt{(\frac{Noise}{SlewRate})^2+(Sample Clock Jitter)^2\; seconds\; rms\; (TIE)} |
Channel-Channel Deskew Range | ±9 x time/div. setting, 100 ms max., each channel |
External Timebase Reference (Input) | 10MHz +/-25ppm at 0 to 10dBm into 50 Ohms |
External Timebase Reference (Output) | 10MHz, 5.0 dBm +/-2.5 dBm, sinewave synchronized to reference being used (internal or external reference) |
|
Acquisition - Analog Channels |
Sample Rate (Single-shot) | 10 GS/s on 8 Ch with Enhanced Sample Rate |
Memory Length (4 Ch / 2 Ch / 1Ch) (Number of Segments) | 50M/100M/200M (65,536) |
Memory Length Options (4 Ch / 2 Ch / 1Ch) (Number of Segments) | WR8KHD-500MPT Option: 125M / 250M / 500M (65,535) WR8KHD-1000MPT Option: 250M / 500M / 1000M (65,535) WR8KHD-2000MPT Option: 500M / 1000M / 2000M (65,535) WR8KHD-5000MPT Option: 1250M / 2500M / 5000M (65,535)
Maximum analysis memory: 500 Mpts per channel |
Intersegment Time | 1.5 µs |
Averaging | Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps (Waveforms of ≤500Mpts) |
Enhanced Resolution (ERES) | From 12.5 to 15 bits vertical resolution |
Envelope (Extrema) | Envelope, floor, or roof for up to 1 million sweeps (Waveforms of ≤500Mpts) |
Interpolation | Linear or Sin x/x (2 pt and 5 pt) (Waveforms of ≤500Mpts) |
|
Vertical, Horizontal, Acquisition - Digital Channels |
Maximum Input Frequency | 500 MHz |
Minimum Detectable Pulse Width | 1 ns |
Input Channels | 16 Digital channels (with WR8KHD-MSO option) |
Input Impedance (Flying Leads) | 100 kΩ || 5 pF |
Input Dynamic Range | ± 20V |
Maximum Input Voltage | ±30V Peak |
Minimum Input Voltage Swing | 400mV |
Threshold Groupings | Pod 2: D15 - D8, Pod 1: D7 - D0 |
Threshold Selections | TTL, ECL, CMOS (2.5 V, 3.3 V, 5 V), PECL, LVDS or User Defined |
Threshold Accuracy | ±(3% of threshold setting + 100mV) |
User Defined Threshold Range | ±10 V in 20 mV steps |
User Defined Hysteresis Range | 100 mV to 1.4 V in 100 mV steps |
Sample Rate | 2.5GS/s |
Record Length | Standard: 50 Mpts Any memory option: 500 Mpts |
Channel-to-Channel Skew | 350 ps |
|
Triggering System |
Max Trigger Frequency (C1-C4, Aux In, SMART Trigger) | 2 GHz |
Modes | Acquisition of ≤500Mpts: Normal, Auto, Single, and Stop
Acquisition of >500Mpts: Single |
Sources | Any input channel, Ext, Ext/10, or Line; slope and level unique to each source (except line trigger) |
Coupling | DC, AC, HFRej, LFRej |
Pre-trigger Delay | 0-100% of memory size |
Post-trigger Delay | No limitation |
Hold-off | From 2 ns up to 20 s or from 1 to 99,999,999 events |
Trigger and Interpolator Jitter | ≤ 2.5 psrms (typical) <0.1 ps RMS (typical, software assisted) |
Internal Trigger Level Range | ±4.1 div from center (typical) |
External Trigger Level Range | Ext (±0.4 V); Ext/10 (±4 V) |
Maximum Trigger Rate | 650,000 waveforms/second (in Sequence Mode, up to 8 channels) |
Trigger Sensitivity with Edge Trigger ProBus Inputs | 2.5 div @ < 1 GHz 2 div @ < 1 GHz 1.5 div @ < 500 MHz 1 div @ < 200 MHz 0.9 div @ < 10 MHz (DC, AC, and LFRej coupling, ≥ 2 mV/div, 50 Ω) |
Trigger Sensitivity with Edge Trigger (Aux Input) | 4.5 div @ < 1 GHz 1.5 div @ < 500 MHz 1 div @ < 200 MHz 0.9 div @ < 10 MHz (DC, AC, and LFRej coupling) |
|
Trigger Types |
Edge | Triggers when signal meets slope (positive, negative, or either) and level condition. |
Width | Triggers on positive or negative glitches with selectable widths. minimum width 750ps, Maximum Width: 20s |
Glitch | Triggers on positive or negative glitches with selectable widths. minimum width 750ps, Maximum Width: 20s |
Window | Triggers when signal exits a window defined by adjustable thresholds |
Pattern | Logic combination (AND, NAND, OR, NOR) of up to 9 inputs (8 channels and external trigger input). Each source can be high, low, or don’t care. The High and Low level can be selected independently. Triggers at start or end of the pattern. |
Runt | Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1 ns and 20 ns. |
Slew Rate | Trigger on edge rates. Select limits for dV, dt, and slope. Select edge limits between 1 ns and 20 ns. |
Interval | Triggers on intervals selectable between 1 ns and 20 s. |
Dropout | Triggers if signal drops out for longer than selected time between 1 ns and 20 s. |
Measurement Trigger | Select from a large number of measurement parameters. Trigger on a measurement value with qualified limits. |
Multi-Stage: Qualified (Timeout or State/Edge Qualified) | Triggers on any input source only if a defined state or edge occurred on another input source. Delay between sources is selectable by time or events. (Note: event B pattern trigger cannot include analog channels). |
Triggers with Exclusion Technology | Glitch, Width, Interval, Runt, Slew Rate - Trigger on intermittent faults by specifying the expected behavior and triggering when that condition is not met |
Multi-Stage: Qualified First | In Sequence acquisition mode, triggers repeatably on event B only if a defined pattern, state, or edge (event A) is satisfied in the first segment of the acquisition. Holdoff between sources is selectable by time or events. (Note: event B pattern trigger cannot include analog channels). |
TriggerScan | A Trigger Trainer analyzes the waveforms, identifies normal behavior, and then sets up a large set of rare event smart trigger setups that target abnormal behavior. The trainer ‘learns’ trigger setups based on slew rates, periods, amplitudes outside of a range and then applies them sequentially. |
|
Low Speed Serial Protocol Trigger (Optional) |
Supported Protocols | I2C, I3C, SPI (SPI, SSPI, SIOP), UART-RS232, CAN1.1, CAN2.0, CAN FD, LIN, FlexRay, SENT, MIL-STD-1553, AudioBus (I2S, LJ, RJ, TDM), USB1.x/2.0, SPMI |
|
Measurement Tools |
Measurement Functionality | Display up to 12 measurement parameters together with statistics, including mean, minimum, maximum, standard deviation, and total number. Each occurrence of each parameter is measured and added to the statistics table. Histicons provide a fast, dynamic view of parameters and waveshape characteristics. Parameter math allows addition, subtraction, multiplication, or division of two different parameters. Parameter gates define the location for measurement on the source waveform. Parameter accept criteria define allowable values based on range setting or waveform state. |
Measurement Parameters - Horizontal + Jitter | Cycles (number of), Delay (from trigger, 50%), Δ Delay (50%), Duty Cycle (50%, @level), Edges (number of, @level), Fall Time (90-10, @levels), Frequency (50%, @level), Half Period (@level), Hold Time (@level), N Cycle Jitter (peak-peak), Number of Points, Period (50%, @level), Δ Period (@level), Phase (@level), Rise Time (10-90, @levels), Setup (@levels), Skew (@levels), Slew Rate (@levels), Time Interval Error (@level), Time (@level), Δ Time (@level), Width (50%, @level), Δ Width (@level), X(value)@max, X(value)@min |
Measurement Parameters - Vertical | Amplitude, Base, Level@X, Maximum, Mean, Median, Minimum, Peak-to-Peak, RMS, Std. Deviation, Top. |
Measurement Parameters - Pulse | Area, Base, Fall Time (90-10, 80-20, @levels), Overshoot (positive, negative), Rise Time (10-90, 80-20, @levels), Top, Width (50%) |
Measurement Parameters - Statistical (on Histograms) | Full Width (@ Half Max, @ %), Amplitude, Base, Peak @ Max Population, Maximum, Mean, Median, Minimum, Mode, Range, RMS, Std. Deviation, Top, X(value)@Peak, Peaks (number of), Percentile, Population (@bin, total) |
|
Math Tools |
Math Functionality | Display up to 12 math functions traces (F1-F12). The easy-to-use graphical interface simplifies setup of up to two operations on each function trace, and function traces can be chained together to perform math-on-math. |
Math Operators - Basic Math | Average (summed), Average (continuous), Difference (-), Envelope, Floor, Invert (negate), Product (x), Ratio (/), Reciprocal, Rescale (with units), Roof, Sum (+). |
Math Operators - Digital (incl. with MSO models/options) | Digital AND, Digital DFlipFlop, Digital NAND, Digital NOR, Digital NOT, Digital OR, Digital XOR |
Math Operators - Filters | Enhanced resolution (to 15 bits vertical), Interpolate (cubic, quadratic, sinx/x), (SinX)/x. |
Math Operators - Frequency Analysis | FFT (power spectrum, magnitude, phase, power density, real, imaginary, magnitude squared), up to full record length. Select from Rectangular, VonHann, Hamming, FlatTop and Blackman Harris windows. |
Math Operators - Functions | Absolute value, Correlation (two waveforms), Derivative, Deskew (resample), Exp (base e), Exp (base 10), Integral, Invert (negate), Log (base e), Log (base 10), Reciprocal, Rescale (with units), Square, Square root, Zoom (identity). |
Math Operators - Other | Segment, Sparse |
|
Measurement and Math Integration |
Measurement and Math Integration | Histograms to display statistical distributions of up to 2 billion measurement parameters. Trend (datalog) of up to 1 million measurement parameters. Track (display parameter vs. time, time-correlated to acquisitions) any parameter. Persistence histogram and persistence trace (mean, range, sigma) |
|
Pass/Fail Testing |
Pass/Fail Testing | Display up to 12 pass/fail queries using a Single or Dual Parameter Comparison (compare All values, or Any value <. ≤, =, >, ≥, within limit ±Δ value or %) or Mask Test (pre-defined or user-defined mask, waveform All In, All Out, Any In, or Any Out conditions). Combine queries into a boolean expression to Pass or Fail IF All True, All False, Any True, Any False, or groups or All or Any, with following THEN Save (waveforms), Stop, Alarm, (send) Pulse, Hardcopy (send email, save screen image, save to clipboard, send to printer), or (save) LabNotebook. |
|
Display System |
Size | Color 15.6 widescreen capacitive touch screen |
Resolution | 1920 x 1080 pixels |
Number of traces | Display a maximum of 40 traces. Simultaneously display channel, zoom, memory, and math. |
Grid Styles | Auto, Single, Dual, Triplex, Quad, Octal, X-Y, Single+X-Y, Dual+X-Y, Tandem, Triad, Quattro, Twelve, Sixteen, Twenty. Supports Normal Display Mode (1 grid style, selectable) or Q-Scape Display Mode (4 different tabs, each with individually selectable grid styles). Q-Scape tabbed displays may be viewed in Single, Dual, or Mosaic mode. |
Waveform Representation | Sample dots joined, or sample dots only |
|
Persistence Display |
Analog and Color-Graded Persistence | Waveforms of ≤500Mpts: Variable saturation levels; stores each trace’s persistence data in memory |
Persistence Types | Waveforms of ≤500Mpts: Select analog, color, or three-dimensional |
Trace Selection | Waveforms of ≤500Mpts: Activate persistence on all or any combination of traces |
Persistence Aging | Waveforms of ≤500Mpts: Select from 500 ms to infinity |
Sweep Display Modes | Waveforms of ≤500Mpts: All accumulated, or all accumulated with last trace highlighted |
|
Zoom Expansion Traces |
Zoom Expansion Traces | Display up to 12 Zoom and 12 Math/Zoom traces |
|
Internal Waveform Memory |
Internal Waveform Memory | 12 active waveform memory traces (M1-M12) store 16 bit/point full length waveforms. Waveforms can be stored to any number of files limited only by the data storage media capacity. |
|
Setup Storage |
Front Panel and Instrument Status | Store to the internal hard drive, over the network, or to a USB-connected peripheral device. |
|
Automatic Setup |
Auto Setup | Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals |
Find Vertical Scale | Automatically sets the vertical sensitivity and offset for the selected channel to display a waveform with the maximum dynamic range |
|
Calibration |
Calibration Stability | Calibration ensured within +/-5°C operating temperature. Manual calibration may be invoked if outside this temperature range. |
|
Processor/CPU |
Type | Intel® Core i5-6500 Quad Core, 3.2 GHz (or better) |
Processor Memory | 16 GB standard for standard memory with 32 GB optional. 32 GB standard for any memory option. |
Operating System | Microsoft Windows® 10 |
Oscilloscope Operating Software | Teledyne LeCroy MAUI™ with OneTouch |
Real Time Clock | Date and time displayed with waveform in hardcopy files. SNTP support to synchronize to precision internal clocks. |
|
Connectivity |
Ethernet Port | 2 x 10/100/1000BaseT Ethernet interface (RJ45 port) |
USB Host Ports | 4 side USB 3.1 Gen1 ports, 2 front USB 3.1 Gen1 ports |
USB Device Port | 1 port - USBTMC over USB 3.1 Gen1 |
GPIB Port (optional) | Supports IEEE - 488.2 (External) |
External Monitor Port | 1 x DisplayPort, supports up to 4096x2304 @ 24 Hz 1 x HDMI, supports up to 4096x2304 @ 60 Hz |
Remote Control | Via Windows Automation, or via LeCroy Remote Command Set |
Network Communication Standard | VXI-11 or VICP, LXI Compatible |
|
Auxiliary Input |
Signal Types | For external trigger input |
Coupling | 50 Ω: DC; 1 MΩ: AC, DC, GND |
Max. Input Voltage | 50 Ω 5V RMS, +/- 10V Peak
1 MΩ 400 V max (Peak DC+AC) |
|
Auxiliary Output |
Signal Types | Select from output signal, control signals, off |
Output Signal | 500 Hz-5 MHz square wave or DC level; 3.3 V +/- 10% into 1 MΩ |
Control Signals | Trigger enabled, trigger out, pass/fail status, off |
Connector Type | BNC, located on rear panel |
|
Probes |
Probing System | ProBus. Automatically detects and supports a variety of compatible probes |
Probes
| Qty. (4) ÷10 Passive Probes |
Scale Factors | Automatically or manually selected depending on probe used |
Calibration Output | Default is 1kHz square wave, 1Vp-p (typical), output to probe hook. Settable from 500 Hz to 1 MHz square wave; 50 mV to 1.0V. |
|
Power Requirements |
Voltage | 100 - 240 VAC ±10% at 47 - 63 Hz 100 - 120 VAC ±10% at 380 - 420 Hz Automatic AC Voltage Selection |
Nominal Power Consumption | 400 W / 400 VA |
Max. Power Consumption | 500 W / 500 VA (with all PC peripherals and active probes connected to 8 channels) |
|
Environmental |
Temperature (Operating) | +5 °C to +40 °C |
Temperature (Non-Operating) | -20 °C to +60 °C |
Humidity (Operating) | 5% to 90% relative humidity (non-condensing) up to +31 °C. Upper limit derates to 50% relative humidity (non-condensing) at +40 °C. |
Humidity (Non-Operating) | 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F |
Altitude (Operating) | Up to 10,000 ft. (3048 m) at or below +30 °C |
Altitude (Non-Operating) | Up to 40,000 ft. (12,192 m) |
Random Vibration (Operating) | 0.31 grms 5 Hz to 500 Hz, 20 minutes in each of three orthogonal axes |
Random Vibration (Non-Operating) | 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes |
Functional Shock | 30 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total |
|
Size and Weight |
Dimensions (HWD) | 13.6”H x 17.5”W x 7.7”D (345 mm x 445 mm x 196 mm) |
Weight | 24.4 lbs (11.1 kg) |
Shipping Weight | 37 lbs (16.8 kg) |
|
Certifications |
CE Certification UL and cUL Listing | CE Compliant, UL and cUL listed; conforms to UL 61010-1 (3rd Edition), UL 61010-2-030 (1st Edition) CAN/CSA C22.2 No. 61010-1-12 |
|
Warranty and Service |
Warranty and Service | 3-year warranty; calibration recommended annually. Optional service programs include extended warranty, upgrades, and calibration services. |
|