Vertical - Analog Channels |
Analog Bandwidth (Max) | 1 GHz |
Bandwidth | 1 GHz |
Analog Bandwidth @ 50 Ω (-3 dB) (ProBus Input) | 1 GHz (≥2 mV/div) |
Analog Bandwidth @ 1 MΩ (-3 dB) (ProBus Input) | 500 MHz (typical) |
Rise Time (10-90%, 50 Ω) | 415 ps (typical) |
Rise Time (20-80%, 50 Ω) | 290 ps (typical) |
Input Channels | 4 |
Vertical Resolution | 8 bits; up to 11 bits with enhanced resolution (ERES) |
Effective Number of Bits (ENOB) | 6.9 |
Sensitivity | 50 Ω: 1 mV-1 V/div, fully variable; 1 MΩ: 1 mV-10 V/div, fully variable |
DC Vertical Gain Accuracy (Gain Component of DC Accuracy) | ±(1%) FS, offset at 0 V |
Channel-Channel Isolation | DC up to rated BW: 40 dB (>100:1), (For any two ProBus input channels, same v/div settings, typical) |
Offset Range | 50 Ω ±1.6 V @ 1 mV - 4.95 mV/div ±4 V @ 5 mV - 9.9 mV/div ±8 V @ 10 mV - 19.8 mV/div ±10 V @ 20 mV - 1 V/div
1 MΩ ±1.6 V @ 1 mV - 4.95 mV/div ±4 V @ 5 mV - 9.9 mV/div ±8 V @ 10 mV - 19.8 mV/div ±16 V @ 20 mV - 100 mV/div ±80 V @ 102 mV - 1.0 V/div ±160 V @ 1.02 V - 10 V/div |
DC Vertical Offset Accuracy | ±(1.5% of offset setting +1% of full scale + 1 mV) (test limit) |
Maximum Input Voltage | 50 Ω: 5V RMS ± 10V peak 1 MΩ: 400 V max. (DC + peak AC < 10 kHz) |
Input Coupling | 1 MΩ: AC, DC, GND; 50 Ω: DC, GND |
Input Impedance | 50 Ω+/-2% or 1 MΩ||17pF, 10 MΩ || 9.5 pF with supplied Probe |
Bandwidth Limiters | 20MHz, 200MHz |
Rescaling | Length: meters, inches, feet, yards, miles; Mass: grams, slugs; Temperature: celsius, fahrenheit, kelvin; Angle: radian, arcdegr, arcmin, arcsec, cycles, revolutions, turns; Velocity: m/s, in/s, ft/s, yd/s, miles/s; Acceleration: m/s2, in/s2, ft/s2, g0; Volume: liters, cubic meters, cubic inches, cubic feet, cubic yards; Force (Weight): newton, grain, ounce, pound; Pressure: pascal, bar, atmosphere (technical), atmosphere (standard), torr, psi; Electrical: volts, amps, watts, volt-amperes, volt-amperes reactive, farad, coulomb, ohm, siemen, volt/meter, coulomb/m2, farad/meter, siemen/meter, power factor; Magnetic: weber, tesla, henry, amp/meter, henry/meter; Energy: joule, Btu, calorie; Rotating Machine: radian/second, frequency, revolution/second, revolution/minute, N m, lb-ft, lb-in, oz-in, watt, horsepower; Other: %. |
|
Horizontal - Analog Channels |
Timebases | Internal timebase common to 4 input channels; an external clock may be applied at the EXT input |
Acquisition Modes | Real-time, Roll, Random Interleaved Sampling (RIS), Sequence |
Time/Division Range | 20 ps/div - 1.6 ks/div RIS available at ≤ 10 ns/div; Roll Mode available at ≥ 100 ms/div and ≤ 5 MS/s |
Clock Accuracy | <=1.5 ppm +(aging of 0.5 ppm/yr from last calibration) |
Sample Clock Jitter | up to 10µs Acquired Time Range: 100fsrms (Internal Timebase Reference)
up to 10ms Acquired Time Range: 360fsrms (Internal Timebase Reference) |
Delta Time Measurement Accuracy | \sqrt{2}*\sqrt{(\frac{Noise}{SlewRate})^2+(Sample Clock Jitter)^2) (RMS)+(clock accuracy*reading)(seconds) |
Jitter Measurement Floor | \sqrt{(\frac{Noise}{SlewRate})^2+(Sample Clock Jitter)^2\; seconds\; rms\; (TIE)} |
Channel-Channel Deskew Range | ±9 x time/div. setting, 100 ms max., each channel |
External Timebase Reference (Input) | 10MHz +/-25ppm into 50 ohms |
External Timebase Reference (Output) | 10MHz, 3.0 dBm +/-1 dBm, sinewave synchronized to reference being used (internal or external reference) |
External Clock | DC to 100 MHz; (50 Ω/1 MΩ), EXT BNC input, Minimum rise time and amplitude requirements apply at low frequencies. |
|
Acquisition - Analog Channels |
Sample Rate (Single-shot) | 10 GS/s on 4Ch 20 GS/s on 2 Ch |
Sample Rate (Repetitive) | 200 GS/s for repetitive signals (20 ps/div to 10 ns/div) |
Memory Length (4 Ch / 2 Ch / 1Ch) (Number of Segments) | 16M / 32M / 32M (5,000) |
Intersegment Time | 1 µs |
Averaging | Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps |
Enhanced Resolution (ERES) | From 8.5 to 11 bits vertical resolution |
Envelope (Extrema) | Envelope, floor, or roof for up to 1 million sweeps |
Interpolation | Linear or Sin x/x (2 pt and 5 pt) |
|
Triggering System |
Max Trigger Frequency (C1-C4, Aux In, SMART Trigger) | 1.0 GHz @ ≥ 10 mV/div (minimum triggerable width 750 ps) |
Modes | Normal, Auto, Single, and Stop |
Sources | Any input channel, Ext, Ext/10, Line or Fast Edge; slope and level unique to each source (except line trigger) |
Coupling | DC, AC, HFRej, LFRej |
Pre-trigger Delay | 0-100% of memory size (adjustable in 1% increments of 100 ns) |
Post-trigger Delay | 0-10,000 Divisions in real time mode, limited at slower time/div settings or in roll mode |
Hold-off | From 2 ns up to 20 s or from 1 to 99,999,999 events |
Trigger and Interpolator Jitter | ≤ 3.5 ps RMS (typical) <0.1 ps RMS (typical, software assisted) |
Internal Trigger Level Range | ±4.1 div from center (typical) |
External Trigger Level Range | Ext (±0.4 V); Ext/10 (±4 V) |
Maximum Trigger Rate | 1,000,000 waveforms/second (in Sequence Mode, up to 4 channels) |
Trigger Sensitivity with Edge Trigger ProBus Inputs | 2 div @ < 1 GHz 1.5 div @ < 500 MHz 1 div @ < 200 MHz 0.9 div @ < 10 MHz (DC, AC, and LFRej coupling, ≥ 10 mV/div, 50 Ω) |
Trigger Sensitivity with Edge Trigger (Aux Input) | 2 div @ <1 GHz 1.5 div @ < 500 MHz 1 div @ < 200 MHz 0.9 div @ < 10 MHz (DC, AC, and LFRej coupling) |
|
Trigger Types |
Edge | Triggers when signal meets slope (positive, negative, or either) and level condition. |
Width | Triggers on positive or negative glitches with selectable widths. Minimum width 500 ps, maximum width: 20 s |
Glitch | Triggers on positive or negative glitches with selectable widths. Minimum width 200 ps, maximum width: 20 s |
Window | Triggers when signal exits a window defined by adjustable thresholds |
Pattern | Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input). Each source can be high, low, or don’t care. The High and Low level can be selected independently. Triggers at start or end of the pattern. |
TV-Composite Video | Triggers NTSC or PAL with selectable line and field; HDTV (720p, 1080i, 1080p) with selectable frame rate (50 or 60 Hz) and Line; or CUSTOM with selectable Fields (1-8), Lines (up to 2000), Frame Rates (25, 30, 50, or 60 Hz), Interlacing (1:1, 2:1, 4:1, 8:1), or Synch Pulse Slope (Positive or Negative). |
Runt | Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1 ns and 20 ns. |
Slew Rate | Trigger on edge rates. Select limits for dV, dt, and slope. Select edge limits between 1 ns and 20 ns. |
Interval | Triggers on intervals selectable between 1 ns and 20 s. |
Dropout | Triggers if signal drops out for longer than selected time between 1 ns and 20 s. |
Measurement Trigger | Select from a large number of measurement parameters trigger on a measurement value with qualified limits. Can be used as only trigger or last event in a Cascade Trigger. |
Multi-Stage: Qualified (Timeout or State/Edge Qualified) | Triggers on any input source only if a defined state or edge occurred on another input source. Delay between sources is selectable by time or events. |
Triggers with Exclusion Technology | Trigger on intermittent faults by specifying the expected behavior and triggering when that condition is not met |
Multi-Stage: Qualified First | In Sequence acquisition mode, triggers repeatably on event B only if a defined pattern, state, or edge (event A) is satisfied in the first segment of the acquisition. Holdoff between sources is selectable by time or events. |
Multi-Stage: Cascade (Sequence) Trigger, Capability | Arm on A event, then Trigger on B event. Or Arm on A event, then Qualify on B event, and Trigger on C event. |
Multi-Stage: Cascade (Sequence) Trigger, Types | Cascade A then B: Edge, Window, Pattern (Logic) Width, Glitch, Interval, Dropout, or Measurement. Measurement can be on Stage B only. Cascade A then B then C (Measurement): Edge, Window, Pattern (Logic), Width, Glitch, Interval, Dropout, or Measurement. Measurement can be on Stage C only. Cascade A then B then C: Edge, Window, Pattern (Logic) |
Multi-Stage: Cascade (Sequence) Trigger, Holdoff | Holdoff between A and B or B and C is selectable by time or number of events. Measurement trigger selection as the last stage in a Cascade precludes a holdoff setting between the prior stage and the last stage. |
TriggerScan | A Trigger Trainer analyzes the waveforms, identifies normal behavior, and then sets up a large set of rare event smart trigger setups that target abnormal behavior. The trainer ‘learns’ trigger setups based on slew rates, periods, amplitudes outside of a range and then applies them sequentially. |
|
Low Speed Serial Protocol Trigger (Optional) |
Supported Protocols | I2C, SPI (SPI, SSPI, SIOP), UART-RS232, CAN1.1, CAN2.0, CAN FD, LIN, FlexRay, MIL-STD-1553, AudioBus (I2S, LJ, RJ, TDM), USB1.x/2.0, SATA |
|
High Speed Serial Protocol Trigger |
Data Rates | (Option WR8K-80B-8B10B TD) 150 Mb/s - 3.125 Gb/s |
Pattern Length | 80 bits, NRZ or 8b10b |
Clock Recovery Jitter | 1 ps rms + 0.3% Unit Interval rms for PRBS data patterns with 50% transition density |
Hardware Clock Recovery Loop BW | PLL Loop BW = Fbaud/5500, 100 Mb/s to 2.488 Gb/s (typical) |
|
Measurement Tools |
Measurement Functionality | Display up to 8 measurement parameters together with statistics, including mean, minimum, maximum, standard deviation, and total number. Each occurrence of each parameter is measured and added to the statistics table. Histicons provide a fast, dynamic view of parameters and waveshape characteristics. Parameter math allows addition, subtraction, multiplication, or division of two different parameters. Parameter gates define the location for measurement on the source waveform. Parameter accept criteria define allowable values based on range setting or waveform state. |
Measurement Parameters - Horizontal + Jitter | Cycles (number of), Delay (from trigger, 50%), Δ Delay (50%), Duty Cycle (50%, @level), Edges (number of, @level), Fall Time (90-10, @levels), Frequency (50%, @level), Half Period (@level), Hold Time (@level), N Cycle Jitter (peak-peak), Number of Points, Period (50%, @level), Δ Period (@level), Phase (@level), Rise Time (10-90, @levels), Setup (@levels), Skew (@levels), Slew Rate (@levels), Time Interval Error (@level), Time (@level), Δ Time (@level), Width (50%, @level), Δ Width (@level), X(value)@max, X(value)@min |
Measurement Parameters - Vertical | Amplitude, Base, Level@X, Maximum, Mean, Median, Minimum, Peak-to-Peak, RMS, Std. Deviation, Top. |
Measurement Parameters - Pulse | Area, Base, Fall Time (90-10, 80-20, @levels), Overshoot (positive, negative), Rise Time (10-90, 80-20, @levels), Top, Width (50%) |
Measurement Parameters - Statistical (on Histograms) | Full Width (@ Half Max, @ %), Amplitude, Base, Peak @ Max Population, Maximum, Mean, Median, Minimum, Mode, Range, RMS, Std. Deviation, Top, X(value)@Peak, Peaks (number of), Percentile, Population (@bin, total) |
|
Math Tools |
Math Functionality | Display up to 8 math functions traces (F1-F8). The easy-to-use graphical interface simplifies setup of up to two operations on each function trace, and function traces can be chained together to perform math-on-math. |
Math Operators - Basic Math | Average (summed), Average (continuous), Difference (-), Envelope, Floor, Invert (negate), Product (x), Ratio (/), Reciprocal, Rescale (with units), Roof, Sum (+). |
Math Operators - Filters | Enhanced resolution (to 11 bits vertical), Interpolate (cubic, quadratic, sinx/x), (SinX)/x. |
Math Operators - Frequency Analysis | FFT (power spectrum, magnitude, phase, power density, real, imaginary, magnitude squared), up to full record length. Select from Rectangular, VonHann, Hamming, FlatTop and Blackman Harris windows. |
Math Operators - Functions | Absolute value, Correlation (two waveforms), Derivative, Deskew (resample), Exp (base e), Exp (base 10), Integral, Invert (negate), Log (base e), Log (base 10), Reciprocal, Rescale (with units), Square, Square root, Zoom (identity). |
Math Operators - Other | Segment, Sparse |
|
Measurement and Math Integration |
Measurement and Math Integration | Histograms to display statistical distributions of up to 2 billion measurement parameters. Trend (datalog) of up to 1 million measurement parameters. Track (display parameter vs. time, time-correlated to acquisitions) any parameter. Persistence histogram and persistence trace (mean, range, sigma) |
|
Pass/Fail Testing |
Pass/Fail Testing | Display up to 8 pass/fail queries using a Single or Dual Parameter Comparison (compare All values, or Any value <. ≤, =, >, ≥, within limit ±Δ value or %) or Mask Test (pre-defined or user-defined mask, waveform All In, All Out, Any In, or Any Out conditions). Combine queries into a boolean expression to Pass or Fail IF All True, All False, Any True, Any False, or groups or All or Any, with following THEN Save (waveforms), Stop, Alarm, (send) Pulse, Hardcopy (send email, save screen image, save to clipboard, send to printer), or (save) LabNotebook. |
|
Display System |
Size | Color 15.4 widescreen capacitive touch screen |
Resolution | WXGA; 1280 x 800 pixels. |
Number of traces | Display a maximum of 16 traces. Simultaneously display channel, zoom, memory and math traces. |
Grid Styles | Auto, Single, Dual, Quad, Octal, X-Y, Single+X-Y, Dual+X-Y, Tandem, Quatro, Twelve, Sixteen, Triple, Hex |
Waveform Representation | Sample dots joined, or sample dots only |
|
Persistence Display |
Analog and Color-Graded Persistence | Variable saturation levels; stores each trace’s persistence data in memory |
Persistence Types | Select analog, color, or three-dimensional |
Trace Selection | Activate persistence on all or any combination of traces |
Persistence Aging | Select from 500 ms to infinity |
Sweep Display Modes | All accumulated, or all accumulated with last trace highlighted |
|
Zoom Expansion Traces |
Zoom Expansion Traces | Display up to 8 Zoom and 8 Math/Zoom traces |
|
Internal Waveform Memory |
Internal Waveform Memory | 4 active waveform memory traces (M1-M4) store 16 bit/point full length waveforms. Waveforms can be stored to any number of files limited only by the data storage media capacity. |
|
Setup Storage |
Front Panel and Instrument Status | Store to the internal hard drive, over the network, or to a USB-connected peripheral device. |
|
Automatic Setup |
Auto Setup | Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals |
Find Vertical Scale | Automatically sets the vertical sensitivity and offset for the selected channel to display a waveform with the maximum dynamic range |
|
Calibration |
Auto Calibration | Ensures specified DC and timing accuracy is maintained for 1 year minimum. |
|
Processor/CPU |
Type | Intel® i5-6500 Quad Core, 3.2 GHz (or better) |
Processor Memory | 8 GB standard, up to 16 GB optional |
Operating System | Microsoft Windows® 10 |
Oscilloscope Operating Software | Teledyne LeCroy MAUI™ with OneTouch |
Real Time Clock | Date and time displayed with waveform in hardcopy files. SNTP support to synchronize to precision internal clocks. |
|
Connectivity |
Ethernet Port | Supports 10/100/1000BaseT Ethernet interface (RJ45 port) |
USB Host Ports | 4 side USB 3.1 Gen1 ports and 1 front USB 2.0 port support Windows compatible devices |
USB Device Port | 1 port - USBTMC over USB 3.1 Gen1 |
GPIB Port (optional) | Supports IEEE - 488.2 (External) |
External Monitor Port | HDMI 1.4 (Qty. 1) and DisplayPort 1.2 (Qty. 1) to support customer-supplied external monitor. Includes support for extended desktop operation with UHD 3840 x 2160 pixel resolution and split-grid capability on external monitor. Supports touch screen integration of external monitor (Note: external display can not use a Fujitsu touch-screen driver). |
Remote Control | Via Windows Automation, or via LeCroy Remote Command Set |
Network Communication Standard | VXI-11 or VICP, LXI Class C (v1.2) Compliant |
|
Auxiliary Input |
Signal Types | Select External Trigger or External Clock Input on the front panel |
Coupling | 50 Ω: DC; 1 MΩ: AC, DC, GND |
Max. Input Voltage | 50 Ω: 5 Vrms; 1 MΩ: 250 Vmax (DC + Peak AC ≤ 10 kHz) |
|
Auxiliary Output |
Signal Types | Select from control signals or Off |
Output Signal | 500 Hz-1 MHz square wave or DC level; 50 mV-1 V into 1 MΩ |
Control Signals | Trigger enabled, trigger out, pass/fail status, fast edge, off |
Connector Type | BNC, located on front |
|
Probes |
Probing System | ProBus. Automatically detects and supports a variety of compatible probes |
Probes
| Qty. (4) ÷10 Passive Probes |
Scale Factors | Automatically or manually selected depending on probe used |
Calibration Output | Default is 1kHz square wave, 1Vp-p (typical), output to probe hook. Settable from 500 Hz to 1 MHz square wave; 50 mV to 1.0V. |
|
Power Requirements |
Voltage | 100-240 VAC ±10% at 50/60 Hz ±5%; 110-120 VAC ±10% at 400 Hz ±5%; Automatic AC Voltage Selection |
Nominal Power Consumption | 285 W / 285 VA |
Max. Power Consumption | 375 W / 375 VA (with all PC peripherals and active probes connected to 4 channels) |
|
Environmental |
Temperature (Operating) | +5 °C to +40 °C |
Temperature (Non-Operating) | -20 °C to +60 °C |
Humidity (Operating) | 5% to 90% relative humidity (non-condensing) up to +31 °C. Upper limit derates to 50% relative humidity (non-condensing) at +40 °C. |
Humidity (Non-Operating) | 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F |
Altitude (Operating) | Up to 10,000 ft. (3048 m) at or below +30 °C |
Altitude (Non-Operating) | Up to 40,000 ft. (12,192 m) |
Random Vibration (Operating) | 0.31 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes |
Random Vibration (Non-Operating) | 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes |
Functional Shock | 30 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total |
|
Size and Weight |
Dimensions (HWD) | 12.44H x 16.42W x 9.37D (316 x 417 x 238 mm) |
Weight | 22.8 lbs. (10.3 kg) |
Shipping Weight | 37.5 lbs. (17.0 kg) |
|
Certifications |
CE Certification UL and cUL Listing | CE Compliant, UL and cUL listed; conforms to UL 61010-1 (3rd Edition), UL 61010-2-030 (1st Edition) CAN/CSA C22.2 No. 61010-1-12 |
|
Warranty and Service |
Warranty and Service | 3-year warranty; calibration recommended annually. Optional service programs include extended warranty, upgrades, and calibration services. |
|